Abstract
Cleaning validation procedures are carried out in order to assure that residues of cleaning agents are within acceptable limits after the cleaning process. Cleaning agents often consist of a mixture of various surfactants which are in a highly diluted state after the water rinsing procedure has been completed. This makes it difficult to find appropriate analytical methods that are sensitive enough to detect the cleaning agents. In addition, it is advantageous for the analytical methods to be simple to perform and to give results quickly.
In this study, four different analytical methods are compared: visual detection of foam, pH, conductivity measurements, and analysis of total organic carbon (TOC). TOC was used as a reference method when evaluating the other three potential methods. The analyses were performed on different dilutions of the cleaning agents Vips Neutral™ , RBS-25™, Debisan™ and Perform™. The results demonstrated that the most sensitive method for analysis of Vips Neutral™, Debisan™ and Perform™ is visual detection of foam, by which it is possible to detect concentrations of cleaning agents down to 10 μg/mL. RBS-25 was not detected below 200 μg/mL, probably because it is formulated with lowfoaming surfactants. TOC analysis is less sensitive but has the advantage of being a quantitative analysis, while visual detection of foam is a semi-quantitative method. Visual detection of foam is easy to perform, gives a quick result, and requires no expensive instrumentation. The sensitivity of each method was found to be dependent upon the type of cleaning agent that was analyzed.
Footnotes
- Copyright © Parenteral Drug Association. All rights reserved.
PDA members receive access to all articles published in the current year and previous volume year. Institutional subscribers received access to all content. Log in below to receive access to this article if you are either of these.
If you are neither or you are a PDA member trying to access an article outside of your membership license, then you must purchase access to this article (below). If you do not have a username or password for JPST, you will be required to create an account prior to purchasing.
Full issue PDFs are for PDA members only.
Note to pda.org users
The PDA and PDA bookstore websites (www.pda.org and www.pda.org/bookstore) are separate websites from the PDA JPST website. When you first join PDA, your initial UserID and Password are sent to HighWirePress to create your PDA JPST account. Subsequent UserrID and Password changes required at the PDA websites will not pass on to PDA JPST and vice versa. If you forget your PDA JPST UserID and/or Password, you can request help to retrieve UserID and reset Password below.