Abstract
In the pharmaceutical industry, the integrity of sterile filters is critical to ensure sterility of filtered products. Filter integrity is frequently tested by measuring gas diffusion across water-contacting hydrophobic or hydrophilic membranes with the same automated test devices. Constant device accuracy over the whole range of possible operating conditions is an especially important requirement, as set by the GMP regulations for product critical devices.
In this paper, we investigate the accuracy of gas diffusion rate and water intrusion rate estimates provided by a batch-operated and a refilling, continuous-flow commercial automated test device used both for diffusive flow tests and water intrusion tests. Tests were performed on custom-designed model filter systems and full-scale filters over a broad range of gas diffusive flow rates and upstream gas volumes. Neither tested device provided accurate measurements of gas diffusion rate when a small gas diffusion flow was measured out of a very large upstream volume. The batch-operated device provided measurements of gas diffusion rates (either gas diffusion or water intrusion rate) with an accuracy that strongly depends on the gas diffusion rate and on the gas volume upstream from the membrane. Gas diffusion rate measurements were particularly biased in diffusive flow tests of filters with less than 500 mL gas upstream volume. Gas diffusion rates were underestimated by as much as −14.5% in diffusive flow tests and −25 % in water intrusion tests. The refilling, continuous flow device generally provided consistent and accurate gas diffusion rate and water intrusion rate measurements within less than 5% of the reference value, practically independent of the gas diffusion flow rate and upstream volume value. A serious bias was only noted in diffusion flow tests at very high upstream volumes and low gas diffusion rate. The results reported in this paper show the importance of qualifying the automated test devices used to assess sterile filter integrity.
Footnotes
- Copyright © Parenteral Drug Association. All rights reserved.
PDA members receive access to all articles published in the current year and previous volume year. Institutional subscribers received access to all content. Log in below to receive access to this article if you are either of these.
If you are neither or you are a PDA member trying to access an article outside of your membership license, then you must purchase access to this article (below). If you do not have a username or password for JPST, you will be required to create an account prior to purchasing.
Full issue PDFs are for PDA members only.
Note to pda.org users
The PDA and PDA bookstore websites (www.pda.org and www.pda.org/bookstore) are separate websites from the PDA JPST website. When you first join PDA, your initial UserID and Password are sent to HighWirePress to create your PDA JPST account. Subsequent UserrID and Password changes required at the PDA websites will not pass on to PDA JPST and vice versa. If you forget your PDA JPST UserID and/or Password, you can request help to retrieve UserID and reset Password below.