RT Journal Article SR Electronic T1 Cleaning Validation for Residual Estimation of Mometasone Furoate on Stainless-Steel Surface of Pharmaceutical Manufacturing Equipment Using a UHPLC-UV Method JF PDA Journal of Pharmaceutical Science and Technology JO PDA J Pharm Sci Technol FD Parenteral Drug Association (PDA) SP 41 OP 48 DO 10.5731/pdajpst.2018.009712 VO 74 IS 1 A1 Coelho, Ana Silva A1 Arribada, Raquel Gregorio A1 Lages, Eduardo Burgarelli YR 2020 UL http://journal.pda.org/content/74/1/41.abstract AB Cleaning validation is the documented evidence that shows the effectiveness of cleaning procedures for the removal of product residues and other contaminants. The cleaning procedures must be validated and methods to determine trace amounts of drugs have to be considered with special attention. An ultra-high-performance liquid chromatography-ultraviolet (UHPLC-UV) method for the determination of mometasone furoate residues on stainless-steel surfaces was developed and validated in order to control a cleaning procedure. The chromatography separation was achieved on a Waters Acquity UPLC HSS T3 column (50 × 2.1 mm, 1.8 µm) at 40°C using acetonitrile and water (1:1, v/v) as the mobile phase at a flow rate of 0.5 mL/min. The injection volume was 2 µL, and the detection was performed at 254 nm. The swab and rinse procedures were optimized in order to obtain a recovery higher than 90% of mometasone furoate from stainless-steel surfaces, using ethanol as the extraction solvent. The method was validated in the range of 0.2–2.6 µg/mL and showed appropriate selectivity, limit of detection and quantification, linearity, precision, accuracy, and robustness. This method was found to be simple, fast, and sensitive for determination of mometasone furoate residues and, therefore, can be used for cleaning validation analysis.