PT - JOURNAL ARTICLE AU - Subrata Chakraborty AU - Vaidehi Soman AU - Vinayak Dhumal TI - Alternate Approach for Defining Worst Case in Steam Sterilization Validation of Porous/Hard Goods Loads AID - 10.5731/pdajpst.2019.011304 DP - 2021 Jan 01 TA - PDA Journal of Pharmaceutical Science and Technology PG - 91--99 VI - 75 IP - 1 4099 - http://journal.pda.org/content/75/1/91.short 4100 - http://journal.pda.org/content/75/1/91.full SO - PDA J Pharm Sci Technol2021 Jan 01; 75 AB - For steam sterilization load validation of porous/hard goods loads, regulatory guidelines state to identify and validate the worst-case challenge that is the most difficult to sterilize. The maximum load and the minimum load are usually considered as worst-case challenges, and both are validated. This article demonstrated with examples how the control strategy determined the worst-case load in the steam sterilizer. It further established with examples how a suitably designed control strategy could nullify the impact of varying load characteristics (e.g., mass, configuration, and air removal challenges) in porous/hard goods loads and make it load size independent.