RT Journal Article SR Electronic T1 Case Study: Compatibility Testing of Pre-Filled Syringe Stopper Technology with Vaporized Hydrogen Peroxide Terminal Sterilization Process JF PDA Journal of Pharmaceutical Science and Technology JO PDA J Pharm Sci Technol FD Parenteral Drug Association (PDA) SP 512 OP 513 DO 10.5731/pdajpst.2024.012965 VO 78 IS 4 A1 Mattila, Juha A1 Kienzle, Michael A1 Ruley, Arja YR 2024 UL http://journal.pda.org/content/78/4/512.abstract AB STERIS and W.L. GORE collaborated on a case study testing the compatibility of a new prefilled syringe plunger design with VHP terminal sterilization. VHP chamber conditions require deep vacuum pulsing, which may represent challenges to prefilled syringe container integrity. The growing industry trend toward VHP sterilization is driven by the FDA search for alternative sterilization methods to EO and the recent publication of a VHP specific process standard. The purpose of the study is to test and report compatibility of the new 0.5 mL GORE IMPROJECT plunger, a silicone free syringe solution for ophthalmic application, with VHP sterilization. Various challenges have been reported when using conventional, siliconized, prefilled syringe systems for intravitreal injections such as subvisible particles, inflammation, silicone floaters, and intraocular pressure increases. The GORE plunger eliminates the need for silicone oil as a lubricant on the plunger and barrel, while meeting strict container closure and terminal sterilization requirements of ophthalmic applications. This case study presents successful results of deep vacuum VHP terminal sterilization process compatibility with the GORE plunger design and material composition. Test results include primary container integrity, stopper off-gassing/ingress, and visual inspection. Principles of VHP vacuum sterilization process, test cycle configuration, and its main parameters are presented.