PT - JOURNAL ARTICLE AU - Twohy, Christine W. AU - Duran, Anthony P. AU - Peeler, James T. TI - Extraction of Bacterial Endotoxin from Medical Devices DP - 1986 Nov 01 TA - PDA Journal of Pharmaceutical Science and Technology PG - 287--291 VI - 40 IP - 6 4099 - http://journal.pda.org/content/40/6/287.short 4100 - http://journal.pda.org/content/40/6/287.full SO - PDA J Pharm Sci Technol1986 Nov 01; 40 AB - A 1984 collaborative study was undertaken to determine the efficacy of an extraction procedure for bacterial endotoxin. Twelve laboratories from government and industry participated in the study. The study consisted of three samples of 8 medical devices contaminated with duplicate amounts at various endotoxin concentrations derived from three different sources: a raw Pseudomonas cepacia (ATCC #25416) endotoxin, the USP reference standard endotoxin EC-5, and Difco E. coli endotoxin 055.B5 (lot 504089). The extraction procedure was performed on the devices and the eluates were tested for endotoxin content using the Limulus Amebocyte Lysate gel tube test. The percent recovery of endotoxin from the devices ranged from 42 to 120%.