PT - JOURNAL ARTICLE AU - Adner, Niklas AU - Flinik, Ola AU - Nystrand, Rolf AU - Pihl, Berthil AU - Vegis, Peter TI - Collaborative Study of a Control Standard Endotoxin Using the Limulus Amebocyte (Gel-Clot) Test DP - 1991 Mar 01 TA - PDA Journal of Pharmaceutical Science and Technology PG - 88--93 VI - 45 IP - 2 4099 - http://journal.pda.org/content/45/2/88.short 4100 - http://journal.pda.org/content/45/2/88.full SO - PDA J Pharm Sci Technol1991 Mar 01; 45 AB - A collaborative study was undertaken to determine the potency in endotoxin Units (EU) and International Units (IU) of a control standard endotoxin, LIF-1. Five laboratories from the Swedish Pharmaceutical Industry participated in the study. As reference preparations, two official standards, USP reference standard endotoxin, EC-5 (expressed in EU) and WHOs international standard endotoxin (expressed in IU),were used. The study was performed using the Limulus Amebocyte Lysate (LAL), gel-clot test. The test protocol included dilutions of the endotoxin in steps of 1:1.25 instead of the conventional 1:2 step dilution method. This gave more precise and standardized results. The content of one vial of LIF-1 (CSE) was calculated to be 22 EU and 16IUwhich indicates that I EU (USP) corresponds to 0.7 IU (WHO).