RT Journal Article SR Electronic T1 Ethylene Oxide Sterilization: III—Influence of Carrier Nature in a Biological Monitor Performance JF PDA Journal of Pharmaceutical Science and Technology JO PDA J Pharm Sci Technol FD Parenteral Drug Association (PDA) SP 155 OP 158 VO 48 IS 3 A1 Pinto, Terezinha J. A. A1 Saito, Takako A1 Iossif, Mario YR 1994 UL http://journal.pda.org/content/48/3/155.abstract AB The ethylene oxide sterilization process control should take into account several parameters, so that only the biological monitor can integrate in all of them, presenting a biological challenge similar to the product under sterilization conditions. The authors evaluated spore resistance on several different carriers: paper, aluminum foil and plastic. The influence was studied through the challenge of sub-lethal cycles, followed by the lethality study and the D-value calculation. There was a significant statistical difference in spore resistance between paper and plastic carriers.