RT Journal Article SR Electronic T1 Microbiological Evaluation of Reused Catheter Guides in a Brazilian Hospital JF PDA Journal of Pharmaceutical Science and Technology JO PDA J Pharm Sci Technol FD Parenteral Drug Association (PDA) SP 356 OP 365 VO 60 IS 6 A1 Mônica Valero Da Silva A1 Marina Baquerizo Martinez A1 Terezinha De Jesus Andreoli Pinto YR 2006 UL http://journal.pda.org/content/60/6/356.abstract AB Within Brazilian hospitals, as well as in hospitals in other countries, the safety of reusing single-use medical devices is being questioned. At the same time, there has been a constant, growing increase in the number of such reprocessed products. In an attempt to evaluate such procedures, we analyzed 30 catheter guide units that were reused four times in patients at a public hospital of the state of São Paulo, Brazil. After each use, the catheter guides had been sterilized with a mixture of ethylene oxide/chlorofluorocarbons (12:88). Each unit analyzed was provided by the hospital individually packed and ready for a possible new use. For each catheter guide, a series of measurements were taken and tests run: microbial counts (pour plate), direct inoculation sterility test, bacterial endotoxin, in vitro cytotoxicity, physical evaluation by scanning electron microscopy, and microbial identification via biochemical assays. The results confirmed the presence of bacteria considered pathogenic to immunologicallycompromised patients with a maximum limit of 104 cfu/unit (catheter guide). Furthermore, bacterial endotoxins and significant modifications of the catheter guides' physical structure were also detected. Thus, the common practice of reusing single-use devices may increase patients' risk of infection or pyrogenic reactions, adding to the total period of hospitalization.