TY - JOUR T1 - Detection of Silicone Oil Leakages in Freeze Dryers JF - PDA Journal of Pharmaceutical Science and Technology JO - PDA J Pharm Sci Technol SP - 481 LP - 485 DO - 10.5731/pdajpst.2011.00748 VL - 65 IS - 5 AU - U. Meissner AU - Harald Stahl AU - D. Steinkellner Y1 - 2011/09/01 UR - http://journal.pda.org/content/65/5/481.abstract N2 - As a small leakage from the silicone oil tubes in pharmaceutical freeze dryers may go undetected for several cycles, it provides a major threat to both product safety and the economic performance of such installations. A properly mounted mass spectrometer offers a non-invasive method of detecting the smallest amounts of silicone oil down to 1 ppm. Additionally the system will improve leak testing procedures and can also be used as a process analytical technology tool to improve process control. LAY ABSTRACT: Drugs, and in particular parenterals, need to have a very high level of purity. A production process such as freeze drying carries the inherent risk of contamination by process fluids. This should not happen during routine production, but in the event of a malfunction small amounts of these fluids may contaminate the drug product. The purpose of this paper is to introduce a method which is able to detect such malfunctions before they can develop into a risk for drug purity and ultimately patient safety. ER -