TY - JOUR T1 - Low Leachable Container System Consisting of a Polymer-Based Syringe with Chlorinated Isoprene Isobutene Rubber Plunger Stopper JF - PDA Journal of Pharmaceutical Science and Technology JO - PDA J Pharm Sci Technol SP - 713 LP - 724 DO - 10.5731/pdajpst.2015.01080 VL - 69 IS - 6 AU - Hideaki Kiminami AU - Katsuyuki Takeuchi AU - Koji Nakamura AU - Yoshihiko Abe AU - Philippe Lauwers AU - William Dierick AU - Keisuke Yoshino AU - Shigeru Suzuki Y1 - 2015/11/01 UR - http://journal.pda.org/content/69/6/713.abstract N2 - A 36 month leachable study on water for injection in direct contact within a polymer-based prefillable syringe consisting of a cyclo olefin polymer barrel, a chlorinated isoprene isobutene rubber plunger stopper, a polymer label attached on the barrel, and a secondary packaging was conducted at 25 ± 2 °C and 60 ± 5% relative humidity. Through the various comparison studies, no difference in the leachable amounts was observed between this polymer-based prefilled syringe and a glass bottle as a blank sample reference by 36 months. No influence on the leachables study outcome was noted from the printed label and/or label adhesive or from the secondary packaging. In an additional study, no acrylic acid used as the label adhesive leachable was detected by an extended storage for 45 months at 25 ± 2 °C and 60 ± 5% relative humidity as a worst case. To obtain more details, a comparison extractable study was conducted between a cyclo olefin polymer barrel and a glass barrel. In addition, chlorinated isoprene isobutene rubber and bromo isoprene isobutene rubber were compared. As a result, no remarkable difference was found in the organic extractables for syringe barrels. On the other hand, in the case of element extractable analysis, the values for the cyclo olefin polymer barrel were lower than that for the glass barrel. For the plunger stoppers, the chlorinated isoprene isobutene rubber applied in this study was showing a lower extractable profile as compared to the bromo isoprene isobutene rubber, both for organic and element extractables. In conclusion, the proposed polymer-based prefillable syringe system has great potential and represents a novel alternative that can achieve very low level extractable profiles and can bring additional value to the highly sensitive biotech drug market.LAY ABSTRACT: A 36 month leachable study on water for injection in direct contact within a cyclo olefin polymer barrel and chlorinated isoprene isobutene rubber plunger stopper that has a polymer label attached to the barrel and is wrapped into a secondary packaging was conducted at 25 °C and 60% relative humidity. Through the various comparison studies, no difference in the leachable amounts was observed between polymer-based prefilled syringes and a glass bottle as a blank sample reference by 36 months. No influences on the leachables study outcome were noted from the secondary packaging. To obtain more details, a comparison extractable study was conducted between the cyclo olefin polymer and the glass barrel. In addition, chlorinated isoprene isobutene rubber and bromo isoprene isobutene rubber plunger stoppers were compared as well. As a result, no remarkable difference was found in the organic extractables for barrels. As for element extractable analysis, the values for the cyclo olefin polymer barrel were lower than that for the glass barrel. For the plunger stoppers, the chlorinated isoprene isobutene rubber applied in this study was showing a lower extractable profile as compared to the bromo isoprene isobutene rubber, both for organic and element extractables. In conclusion, the proposed polymer-based prefillable syringe system has great potential and represents a novel alternative that can achieve very low level extractable profiles and can bring additional value to the highly sensitive biotech drug market. ER -