TABLE II

Determination of HUV Based on Linear Regression Model for Vial and Disposable Syringe Configuration

Vial and Disposable Syringe ConfigurationDetermination of HUV from Linear Regression Model
2R vial and 1 mL disposable syringeHUV (mL) = (0.00099*Viscosity) + 0.105
2R vial and 3 mL disposable syringeHUV (mL) = (0.00077*Viscosity) + 0.097
6R vial and 5 mL disposable syringeHUV (mL) = (0.001*Viscosity) + 0.174
10R vial and 10 mL disposable syringeHUV (mL) = (0.00066*Viscosity) + 0.208