Table V

Laplace Pressure Required for Ideal Circular Defects In Water-Filled Vials. The Expected Carbon Dioxide Partial Pressure Is Based on a 20 psi (1034 torr) Overpressure

Defect Size [μm]Laplace Pressure [Torr]Expected Headspace Carbon Dioxide Partial Pressure [Torr]
21080n/a
5432602
10216818