TABLE I

Representative System Performance Data on Flame-Sealed 15 R Clear Tubing Vial Standards. Each Standard Was Measured Five Consecutive Times. The Error Is the Difference Between the Mean and Expected Values. Min and Max Refer to the Minimum and Maximum Values of The Five Consecutive Measurements

LabelTotal Pressure [Torr]Headspace Carbon Dioxide Partial Pressure [Torr]
ExpectedExpectedMeasured
MeanErrorStDevMinMax
LH-15R-2B0.00.00.00.000.000.00.0
LH-15R-2I49.90.00.00.000.010.00.1
LH-15R-2C725.20.00.00.000.000.00.0
LH-15R-2K50.40.10.10.010.010.10.1
LH-15R-2J699.50.70.6−0.110.040.60.6
LH-15R-2L699.91.71.7−0.050.041.71.7
LH-15R-2D724.87.27.80.580.037.87.8
LH-15R-2E725.514.515.40.920.0715.415.5
LH-15R-2F725.229.031.02.010.1330.831.2
LH-15R-2G725.158.061.23.190.1361.061.4
LH-15R-2H725.2145.0151.66.560.04151.5151.6
LH-15R-2A550.2550.2549.6−0.590.15549.4549.8