TABLE III

Flow Effective Diameters Reported by Lenox Laser as Compared to Those Acquired on the Lighthouse LRM System. Lighthouse Values Show the Averages of One Measurement on 10 Vials for Each Nominal Diameter Size

Nominal Diameter [μm]Effective Defect Size [μm]
Lenox LaserLighthouse Instruments
MeanStDevMeanStDev
22.30.622.40.58
55.00.544.80.65
109.21.188.90.73