Headspace Carbon Dioxide Partial Pressure in Sample Vials with Laser-Drilled Defects, Both Before (T0) and After (T1) conducting the CCI Test within ∼24 h of Sample Preparation. Values Show the Averages of Five Consecutive Measurements on Each Vial. All Vials with a Liquid Fill Had the Defect Located Below the liquid Level
Nominal Defect Size [μm] | Fill | Headspace Carbon Dioxide Partial Pressure [Torr] | Number of Leaking Vials Detecteda | |||||||
---|---|---|---|---|---|---|---|---|---|---|
T0 | T1 | |||||||||
Mean | StDev | Min | Max | Mean | StDev | Min | Max | |||
2 | Empty | 1 | 0.0 | 1 | 1 | 96 | 45.3 | 26 | 175 | 10/10 |
PBSb | 0 | 0.1 | 0 | 0 | 0 | 0.1 | 0 | 0 | 0/10 | |
BSAc in PBS | 0 | 0.1 | 0 | 1 | 0 | 0.1 | 0 | 0 | 0/10 | |
5 | Empty | 1 | 0.0 | 1 | 1 | 567 | 33.9 | 498 | 616 | 10/10 |
PBS | 0 | 0.1 | 0 | 0 | 233 | 169.3 | 21 | 476 | 10/10 | |
BSA in PBS | 0 | 0.1 | 0 | 1 | 181 | 122.4 | 25 | 420 | 10/10 | |
10 | Empty | 1 | 0.0 | 1 | 1 | 559 | 21.1 | 520 | 589 | 10/10 |
PBS | 0 | 0.1 | 0 | 0 | 605 | 181.0 | 261 | 778 | 10/10 | |
BSA in PBS | 0 | 0.1 | 0 | 0 | 575 | 115.2 | 471 | 801 | 10/10 | |
NCd | Empty | 1 | 0.0 | 1 | 1 | 1 | 0.0 | 1 | 1 | 0/3 |
GPCe | Empty | 1 | 0.0 | 1 | 2 | 709 | 1.0 | 708 | 710 | 3/3 |