Adequate Sensitivity Ramifications of the UF Adjustment to the AET Applied to a Medical Device. Is Method Sensitivity Adequate to Achieve the Adjusted AET?a
# of Devices Required per Day | AET (µg/mL) | LoD (µg/mL)b | Is the LoD ≤ the Adjusted AET? | |||
---|---|---|---|---|---|---|
No UF | UF = 2 | UF = 5 | UF = 10 | |||
1 | 120 | 0.05 | Yes | Yes | Yes | Yes |
2 | 120 | 0.05 | Yes | Yes | Yes | Yes |
4 | 120 | 0.05 | Yes | Yes | Yes | Yes |
8 | 120 | 0.05 | Yes | Yes | Yes | Yes |
10 | 120 | 0.05 | Yes | Yes | Yes | No |