Table III

Percentage of Compounds Correctly Reported as at or above the Analytical Evaluation Thresholda

DetectorUF = 1UF = 2UF = 3UF = 4
GCMSGCMS-FIDGCMSGCMS-FIDGCMSGCMS-FIDGCMSGCMS-FID
LCMS54%63%85%86%91%92%93%93%
LCMS-UV63%70%90%91%95%95%95%96%
LCMS-CAD62%71%90%91%95%95%96%96%
LCMS-UV-CAD69%75%93%94%97%97%97%97%
  • a Abbreviations: UF is uncertainty factor; GCMS is gas chromatography-mass spectrometry; FID is flame ionization detection; LCMS is liquid chromatography-mass spectrometry; UV is ultraviolet-visible detector; CAD is charged aerosol detector.