Average Probability of Detection for All Types and Sizes of Particles (Black, Clear, and Fiber) as a Function of SUS Component Size (Development Test Kit)
Bag Volume | 0.05 L | 5 L | 50 L |
Bag Surface Area | 71 cm2 | 961 cm2 | 4053 cm2 |
Probability of Detection | 70% | 56% | 46% |
Tubing Line Length | 1050 mm | 4100 mm | |
Probability of Detection | 75% | 21% |