TABLE V

Outside-In He Leak Measurements in One Laboratory

Type of Artificially Created LeakNominal SizeNnAvgMinMaxSpan
Fused silica capillary in viala10 µm× 11 mmb1306.7E-056.2E-057.3E-050.07
Fused silica capillary in viala10 µm× 12 mmb10017.3E-056.4E-058.1E-050.10
Laser-drilled hole in steel tubing in vial5 µm (FED)3518.2E-043.5E-041.1E-030.50
Laser-drilled hole in steel tubing in vial10 µm (FED)3513.3E-032.6E-033.9E-030.18
Laser-drilled hole in steel tubing in vial20 µm (FED)3511.6E-021.4E-021.8E-020.11
  • Notes: FED = flow effective diameter based on air flow measurements (differences up to 20% were reported between certified flow rates). N = Number of prepared units. n = Number of replicate measurements per unit.

  • a Capillary was glued into rubber stopper of a vial.

  • b Nominal internal diameter x length of capillary.