TABLE I

Strategies for Producing Protective Concentrations

Description of LeachableQuantitation StrategyUncertainty Factor (UF)1
Leachables with a confirmed IDUse of compound-specific RRF1.183
Leachables with an unconfirmed IDUse of surrogate-matched RRF2
Leachables with a proposed but unconfirmed IDUse of RRFmedian2 (for GC/MS)2
Unidentified leachableUse of RRFmedian2 (for GC/MS)2
  • GC/MS, gas chromatography with mass spectrometric detection; ID, identification; LC/MS, liquid chromatography with mass spectrometric detection RRF, relative response factor; UF, uncertainty factor

  • 1Reported Concentration = Calculated Concentration × UF.

  • 2Given the much greater RRF variation and more poorly defined RRF distributions for LC/MS, a UF for LC/MS is not proposed, although a minimum value of 5 may be appropriate.

  • 3Although use of such a small UF is consistent with the rigorous application of the concept of protection, use of such a small value may have limited practical value.