TABLE V

Measured Solvent Composition of Ethanol/Water Mixtures After Dialyzer and Tubing Set Extraction

Sample Type10% Ethanol in Water20% Ethanol in Water40% Ethanol in Water70% Ethanol in Water
Hemodialyzer and bloodline tubing set recirculation extraction9.9%18.4%37.4%65.6%
  • Samples were analyzed in triplicate and the average value reported.