TABLE VIII

Recovery in Three Levels of the API A′

Surface Area of Plates Sampled: 40 cm2
Amount of standard 5.70 ppm CTOC-VCSN (ppm C)TOC-VWP (ppm C)HPLC (ppm API A′)
Sample 16.3365.8847.680
Sample 26.4365.8236.410
Sample 36.9166.4557.440
Blanc filtered swab1.3151.245*
High-purity water blank0.2370.287*
5.7 ppmC real6.3065.6949.890
Average6.5636.0547.180
RSD (%)4.7245.7589.359
% recovery86%89%73%
Amount of standard 2.87 ppm CTOC-VCSN (ppm C)TOC-VWP (ppm C)HPLC (ppm API A′)
Sample 12.4842.4411.680
Sample 23.5622.0743.073
Sample 34.0613.9693.332
Blank filtered swab1.3151.245*
High-purity water blanc0.2370.287*
2.87 ppmC real3.3532.4045.059
Average3.3692.8282.694
RSD (%)23.92535.53832.612
%recovery55%75%54%
Amount of standard 1, 43 ppm CTOC-VCSN (ppm C)TOC-VWP (ppm C)HPLC (ppm API A′)
Sample 11.6551.4801.495
Sample 21.5961.4931.606
Sample 31.4251.3711.105
Blanc filtered swab1.3151.245*
High purity water blanc0.1990.239*
1.43 ppmC real1.6981.5272.472
Average1.5591.4481.402
RSD (%)7.6644.62718.320
% recovery16%16%57%
Surface Area Sampled in the Equipment: 400 cm2
Amount of standard 5, 34 ppm CTOC-VCSN (ppm C)TOC-VWP (ppm C)HPLC (ppm API A′)
Sample 15.694.9618.16
Sample 24.744.9914.97
Sample 35.905.3120.34
Blanc filtered swab1.391.30*
High purity water blanc0.230.28*
Average5.445.0917.83
RSD (%)11.4110.7015.12
% recovery75%71%70%
  • * Does not apply.