Bioload on Glove Inner Side; Samples Taken from Existing Production Systems (“After Disinfection” = 1–12 h After Applying Disinfection)
Total Number of Samples | Number of Samples With [CFU/Sample] | Max. Count* | |||
---|---|---|---|---|---|
0 | 1–5 | 6–10 | >10 | ||
103 | Current bioload | ||||
30 | 62 | 9 | 2 | 41 | |
103 | Bioload after disinfection | ||||
89 | 14 | 0 | 0 | 4 | |
53 | Bioload after 11 production batches (=2 weeks) | ||||
9 | 32 | 5 | 7 | 69 |