TABLE I

GC-TEA Conditions for the Final Analytical Method

Gas Chromatograph:HP 5890 Series II Plus (Agilent Technologies)
    A/D Interface:PE Nelson Series 900
    Detector:Thermedics Thermal Energy Analyzer (TEA) model 543
    Carrier:Helium
    Pressure:9.8 psi at 90 °C, Constant Flow
    Column Flow:5 mL/min
    Injector Temperature:280 °C
    Injection Mode:Splitless
    Injection Volume:2 μL
    Liner:4 mm Straight
    Purge Flow:On at 1 min, Purge Flow, 60 mL/minute
    Syringe Size:10 μL
    Column:J&W Scientific DB-624 75 m × 0.53 mm; 3.0 μm
    Guard Column:IP Deact 5 m × 0.53 mm (Restek)
Oven Temperature Program:95 °C, hold 10 min, ramp to 120 °C at 2 °C/minute, ramp to 200 °C at 5 °C/min, and hold for 5 min. Run Time = 43.5 min.
    Interface Temperature:250 °C
    Pyrolyzer Temperature:500 °C
    Detector Attenuation:8
    PMT Voltage:700 V