TABLE II

Average TOC, pH, and Conductivity Results of Water Extractsa

SamplesTOC, ppmpHConductivity, μS/cm
Kleenpak Connectors (set of four)0.095.90.88
Kleenpak Filter Capsule8.924.416.0
  • a The results were corrected with the negative control values except pH, where the pH of the negative control for connectors was 5.8 and that for filter capsule was 5.9.