Table IX

Monitoring of Glove Integrity

ActivityTotal Volume SampledSampling TimeMean ≥0.5-μm Total ParticlesaMean ≥5.0-μm Total ParticlesaMean Viable Particlesa
Monitoring a glove with a 75–100-μm pinhole0.45 m315 min 35 s700
Monitoring a glove with a 200–250-μm pinhole0.22 m37 min 51 s000
Monitoring a glove with a cut fingertip0.30 m310 min 5 s1221b
Monitoring a glove with all fingers cut off and when the glove was removed0.23 m38 min 19 s7517c
  • a Data is presented as the number of particles detected per total volume sampled.

  • b The single viable particle was detected when there was movement of the bare finger through the cut fingertip.

  • c Viable particles were detected only after the operator rubbed his hand on his moustache and reinserted his hand into the isolator.