Defect Type | ID Codea | Leak Test Results Air-Filled Syringes | Visual Inspection Resultsb | |||
---|---|---|---|---|---|---|
dP Pa | Pass/Fail | Inspector 1 | Inspector 2 | Inspector 3 | ||
Controls Tested for Ingress | A6 | 11 | P | N | N | N |
A7 | 10 | P | N | N | N | |
A8 | 12 | P | N | N | N | |
A9 | 9 | P | N | N | N | |
A10 | 9 | P | N | N | N | |
5-μm Hole | 106 | 25 | P | N | N | Y |
107 | 71 | F | N | Y | Y | |
108 | 80 | F | N | Y | Y | |
109 | 43 | F | N | N | N | |
110 | 42 | F | N | N | Y | |
10-μm Hole | 121 | 217 | F | Y | Y | Y |
122 | 177 | F | Y | Y | Y | |
123 | 264 | F | Y | Y | Y | |
124 | 231 | F | N | N | Y | |
125 | 161 | F | N | N | N | |
15-μm Hole | 136 | ABORT | F | N | N | Y |
137 | 344 | F | Y | Y | Y | |
138 | 342 | F | Y | Y | Y | |
139 | 350 | F | Y | Y | Y | |
140 | 281 | F | Y | Y | Y |