ISO Dye Ingress Test versus Vacuum Decay Leak Test—Test Site 1
Defect Type | ID Codea | Leak Test Results Air-Filled Syringes | Visual Inspection Resultsb | |||
---|---|---|---|---|---|---|
dP Pa | Pass/Fail | Inspector 1 | Inspector 2 | Inspector 3 | ||
Controls Tested for Ingress | A16 | 7 | P | N | N | N |
A17 | 6 | P | N | N | N | |
A18 | 7 | P | N | N | N | |
A19 | 6 | P | N | N | N | |
A20 | 7 | P | N | N | N | |
5-μm Hole | 106 | 22 | P | N | N | N |
107 | 66 | F | N | N | Y | |
108 | 79 | F | N | Y | Y | |
109 | 44 | F | N | N | Y | |
110 | 42 | F | N | N | N | |
10-μm Hole | 121 | 205 | F | Y | Y | Y |
122 | 175 | F | Y | Y | Y | |
123 | 260 | F | Y | Y | Y | |
124 | 221 | F | N | N | Y | |
125 | 154 | F | N | N | N | |
15-μm Hole | 136 | 388 | F | Y | Y | Y |
137 | 346 | F | Y | Y | Y | |
138 | 335 | F | Y | Y | Y | |
139 | 337 | F | Y | Y | Y | |
140 | 301 | F | Y | Y | Y |