TABLE VII

Modified USP/Ph.Eur. Dye Ingress Test versus Vacuum Decay Leak Test—Test Site 2

Defect TypeID CodeaLeak Test Results Air-Filled Syringes
Visual Inspection Resultsb
dP PaPass/FailInspector 4Inspector 5Inspector 6
Controls Tested for IngressB68PNNN
B78PNNN
B88PNNN
B98PNNY
B108PNNN
5-μm Hole11164FYYY
11254FNNY
11388FYYY
11456FNNN
11546FNNY
10-μm Hole126192FYYY
127184FYYY
128186FYYY
129301FYYY
130194FYYY
15-μm Hole141352FYYY
142356FYYY
143346FYYY
144445FYYY
145371FYYY
  • a Holed syringes are identical to those used for Part 1, ASTM precision and bias studies.

  • b Y = dye seen, N = No dye seen.