Modified USP/Ph.Eur. Dye Ingress Test versus Vacuum Decay Leak Test—Test Site 2
Defect Type | ID Codea | Leak Test Results Air-Filled Syringes | Visual Inspection Resultsb | |||
---|---|---|---|---|---|---|
dP Pa | Pass/Fail | Inspector 4 | Inspector 5 | Inspector 6 | ||
Controls Tested for Ingress | B6 | 8 | P | N | N | N |
B7 | 8 | P | N | N | N | |
B8 | 8 | P | N | N | N | |
B9 | 8 | P | N | N | Y | |
B10 | 8 | P | N | N | N | |
5-μm Hole | 111 | 64 | F | Y | Y | Y |
112 | 54 | F | N | N | Y | |
113 | 88 | F | Y | Y | Y | |
114 | 56 | F | N | N | N | |
115 | 46 | F | N | N | Y | |
10-μm Hole | 126 | 192 | F | Y | Y | Y |
127 | 184 | F | Y | Y | Y | |
128 | 186 | F | Y | Y | Y | |
129 | 301 | F | Y | Y | Y | |
130 | 194 | F | Y | Y | Y | |
15-μm Hole | 141 | 352 | F | Y | Y | Y |
142 | 356 | F | Y | Y | Y | |
143 | 346 | F | Y | Y | Y | |
144 | 445 | F | Y | Y | Y | |
145 | 371 | F | Y | Y | Y |