Modified ISO Dye Ingress Test versus Vacuum Decay Leak Test—Test Site 3
Defect Type | ID Codea | Leak Test Results | Visual Inspection Resultsb | |||
---|---|---|---|---|---|---|
dP Pa | Pass/Fail | Inspector 7 | Inspector 8 | Inspector 10 | ||
Controls Tested for Ingress | A6 | 9 | P | N | Y | N |
A7 | 9 | P | N | Y | N | |
A8 | 10 | P | N | N | Y | |
A9 | 9 | P | N | Y | Y | |
A10 | 17 | P | Y | N | N | |
5-μm Hole | 101 | 57 | F | Y | Y | Y |
102 | 96 | F | Y | Y | Y | |
103 | 43 | F | Y | Y | Y | |
104 | 41 | F | Y | Y | Y | |
105 | 51 | F | Y | Y | Y | |
10-μm Hole | 116 | ABORT | F | Y | Y | Y |
117 | 191 | F | Y | Y | Y | |
118 | ABORT | F | Y | Y | Y | |
119 | ABORT | F | Y | Y | Y | |
120 | 188 | F | Y | Y | Y | |
15-μm Hole | 131 | ABORT | F | Y | Y | Y |
132 | ABORT | F | Y | Y | Y | |
133 | ABORT | F | Y | Y | Y | |
134 | ABORT | F | Y | Y | Y | |
135 | ABORT | F | Y | Y | Y |