TABLE VIII

Modified ISO Dye Ingress Test versus Vacuum Decay Leak Test—Test Site 3

Defect TypeID CodeaLeak Test Results
Visual Inspection Resultsb
dP PaPass/FailInspector 7Inspector 8Inspector 10
Controls Tested for IngressA69PNYN
A79PNYN
A810PNNY
A99PNYY
A1017PYNN
5-μm Hole10157FYYY
10296FYYY
10343FYYY
10441FYYY
10551FYYY
10-μm Hole116ABORTFYYY
117191FYYY
118ABORTFYYY
119ABORTFYYY
120188FYYY
15-μm Hole131ABORTFYYY
132ABORTFYYY
133ABORTFYYY
134ABORTFYYY
135ABORTFYYY
  • a Holed syringes are identical to those used for Part 1, ASTM precision and bias studies.

  • b Y = dye seen, N = No dye seen.