Vacuum Decay Leak Test Studies Performed in Triplicate at Each Test Site
Study | Leak Test Method (target vacuum) | Negative Controlsa (number tested) | Positive Controlsb (number tested) | NIST Flow Meter (number tests) |
---|---|---|---|---|
1 | Gas leak test (250 mbar) | Empty syringes (15) | NA | 0.05 ccm (5) 0.10 ccm (5) |
2 | Gas leak test (250 mbar) | Empty syringes (15) | Empty syringes 5-μm hole (5) 10-μm hole (5) 15-μm hole (5) | NA |
3 | Liquid leak test (0–1 mbar) | Water-filled syringes (15) | NA | 0.05 ccm (5) 0.10 ccm (5) |
4 | Liquid leak test (0–1 mbar) | Water-filled syringes (15) | Empty syringes 5-μm hole (5) 10-μm hole (5) 15-μm hole (5) | NA |
5 | Liquid leak test (0–1 mbar) | Water-filled syringes (15) | Water-filled syringes 5-μm hole (5) 10-μm hole (5) 15-μm hole (5) | NA |