TABLE III

VeriPac 325/LV Leak Test Parameters

Test ParametersGas Leak TestLiquid Leak TestUnits
Pre-Test Vacuum Flush3020second
Time to reach Vacuum 1 Reference0.203.00second
Time to reach Vacuum 2 Reference0.306.00second
Equalization Time5.000.20second
Test Time10.008.00second
Target or Reserve Vacuum2500–1mbar
Vacuum 1 Reference6005mbar
Vacuum 2 Reference3002mbar
Differential Pressure Reference (dP Ref)No more than 25No more than 25Pascal