TABLE VIII

Study 5. Vacuum Decay Liquid Leak Test at 1 mbar—Water-Filled, Holed Syringe Challenge

Differential Pressure Readings (Pa)
Negative Controlan5 μmn10 μmn15 μmn
Site 1Average ± SD15 ± 2.945122 ± 13415599 ± 015599 ± 015
Site 2Average ± SD12 ± 3.445162 ± 5615599 ± 015599 ± 015
Site 3Average ± SD17 ± 2.744a143 ± 1615487 ± 11715570 ± 5115
Combined SitesAverage ± SD15 ± 3.8134143 ± 8445562 ± 8545589 ± 3245
Maximum Reading25599599599
Minimum Reading663322467
% Passed100000
% Failed (Leaked)0100100100
Average of Reported Lab Averages15.5132.3561.8592.0
Standard Deviation of the Lab Averagesbsx¯4.3NANANA
Repeatability
Standard Deviationsr4.813.213.59.1
Reproducibility
Standard DeviationbsR6.0NANANA
Repeatability Limitr13.436.937.725.4
Reproducibility LimitbR16.7NANANA
  • a One data point was eliminated due to procedural error.

  • b Syringes with hole defects were not tested across sites, therefore reproducibility between labs is not applicable (NA).

  • Note: ABORT test results were assigned differential pressure readings of 599 Pa for analysis purposes

  • SD = standard deviation.