PTI VeriPac 325/LV Vacuum Decay Leak Test Parameters
Test Parameters | Liquid Leak Test | Units |
---|---|---|
Pre-Test Vacuum Flush | 20 | s |
Time to reach Vacuum 1 Reference | 3.00 | s |
Time to reach Vacuum 2 Reference | 6.00 | s |
Equalization Time | 0.20 | s |
Test Time | 8.00 | s |
Target or Reserve Vacuum | 0–1 | mbar |
Vacuum 1 Reference | 5 | mbar |
Vacuum 2 Reference | 2 | mbar |
Differential Pressure Reference (dP Ref) | No more than 25 | Pascal |