TABLE I

PTI VeriPac 325/LV Vacuum Decay Leak Test Parameters

Test ParametersLiquid Leak TestUnits
Pre-Test Vacuum Flush20s
Time to reach Vacuum 1 Reference3.00s
Time to reach Vacuum 2 Reference6.00s
Equalization Time0.20s
Test Time8.00s
Target or Reserve Vacuum0–1mbar
Vacuum 1 Reference5mbar
Vacuum 2 Reference2mbar
Differential Pressure Reference (dP Ref)No more than 25Pascal