TABLE II

Calculation of A0 Value in Moist Heat Disinfection at 103.0 °C for 90 s

Time (s)Measured Temp (°C)10(T−80)/1010(T−80)/10 Multiplied by 10 sA0 Values
Come-up phase068.50.10.7
1073.70.22.3
2078.00.66.4
3082.81.919.2
4086.34.342.7
5090.711.6116.2
6092.819.2191.7
7096.342.7427.0
8098.063.7637.3
90100.7116.21162.0
100102.4173.41734.34340
Holding time phase110104.1258.82588.5
120103.7234.22341.9
130103.3211.92118.8
140103.3211.92118.8
150103.3211.92118.8
160103.3211.92118.8
170103.3211.92118.8
180103.3211.92118.8
190103.3211.92118.819762
Come-down phase20098.570.4704.4
21086.74.747.2
22079.30.98.6
23073.70.22.3
24067.60.10.6763
Estimated A0 value24865
(Reference (22)