Master, Negative, and Positive Control Tests

Package IDDescriptionAbsolute Pressure (mbar)
Evacuation 6.00s Equalization 20.00sEvacuation 7.00s Equalization 20.00s
5-1 to 5-5Masters0.4–0.80.1–0.8
2-1 to 2-10Negative controls, empty packages0.5–1.5Not tested
S-1Closure needle puncture1.2Not tested
S-6Closure flange radial cut0.7Not tested
S-7Closure flange edge removed0.8Not tested
V-1Vial land channel1.20.7
V-2Vial land channel50.0 (cycle aborted 26.61s)145.3
V-3Vial land channel1.30.7
V-4Vial land channel50.2 (cycle aborted 16.77s)150.0 (cycle aborted 17.89s)1
V-5Vial land channel1.90.9
V-6Vial valve channel2.10.9
V-7Vial valve channel1.91.1
V-8Vial valve channel1.91.0
V-9Vial land + valve channel7.66.2
V-10Vial land + valve channel11.614.0
V-11Vial land + valve channel50.2 (cycle aborted 13.04 s)146.2
27-1-OVial 3.38 μm hole, open to vial1.5Not tested
27-3-OVial 2.93 μm hole, open to vial1.2Not tested
27-4-OVial 2.39 μm hole, open to vial1.0Not tested
27-5-OVial 2.67 μm hole, open to vial1.1Not tested
27-11-OVial 4.92 μm hole, open to vial3.0Not tested
27-12-OVial 5.09 μm hole, open to vial2.8Not tested
27-13-OVial 4.92 μm hole, open to vial2.9Not tested
27-14-OVial 4.74 μm hole, open to vial2.2Not tested
27-15-OVial 5.58 μm hole, open to vial3.6Not tested
27-6-OVial 12.85 μm hole, open to vial11.5Not tested
27-7-OVial 12.16 μm hole, open to vial12.0Not tested
27-8-OVial 13.18 μm hole, open to vial11.5Not tested
27-9-OVial 13.18 μm hole, open to vial11.6Not tested
27-10-OVial 14.14 μm hole, open to vial13.7Not tested
  • 1 Absolute pressure reference limit was exceeded and test cycle was aborted at test cycle time indicated.