TABLE IV

Test Method Equalization Time Optimization, 0–20 s IVa. Test Parameter Settings

ParameterSetting
Evacuation6.00 s or 7.00 s
EqualizationVariable
Test0.00 s
Absolute pressure reference limit50 mbar
Differential pressure reference limitN/A1
  • 1 Differential pressure reference limit is irrelevant as test time = 0.00s.