TABLE VII

Test Method Sensitivity Verification Using NIST Airflow Leaks VIIa. Test Parameter Settings

ParameterSetting
Evacuation6.00 s
Equalization20.00 s
Test30.00 s
Absolute pressure reference limit50 mbar
Differential pressure reference limit100 Pa