A 2 × 2 Confusion Matrix Showing Flow Cytometry Method Performance In relation to Standard Plate Method in Parallel Testing of Virus Harvest Fluid Production Scale Samples
Standard Plate Method (Positive Result is ≥5300 CFU/mL) | ||
---|---|---|
Positive | Negative | |
Flow cytometry method (positive result is ≥320 counts/mL) | ||
Positive | 71 | 11 |
Negative | 3 | 1481 |