General Operating Parameters for ICP-AES Analysis
Parameter | Typical Range3 | Aqueous/Saline Extract | Ethanol-containing Extract | Saline/Ethanol Extract |
---|---|---|---|---|
Power (kW) | 1.00–1.50 | 1.25 | 1.50 | 1.30 |
Plasma Flow (L/min) | 15–18.5 | 15 | 18 | 18 |
Auxiliary Flow (L/min) | 1.50–2.25 | 1.50 | 1.50 | 1.50 |
Nebulizer Type | Sea Spray | Sea Spray | Micromist | Sea Spray |
Nebulizer Flow (L/min) | 0.55–0.80 | 0.75 | 0.68 to 0.70 | 0.60 |
Viewing Height (mm) | 7–14 | 10 | 13 | 8 |
Replicate Read Time (s) | >15 | 40 | 40 | 40 |
Instrument Stabilization Time1 (s) | Not Specified | 25 | 25 | 25 |
Sample Uptake Rate (mL/min) | 1.2–1.8 (total flow to the nebulizer) | 1.2–1.8 (total flow to the nebulizer) | 1.2–1.8 (total flow to the nebulizer) | 1.2–1.8 (total flow to the nebulizer) |
Pump Rate, rpm | 8–18 | 15 | 10 | 15 |
Rinse Time(2), sec | Not Specified | 60 | 60 | 60 |
Replicate Readings (per standard/sample) | 3 | 3 | 3 | 3 |
Spray Chamber | NA | Cyclonic Double Pass | Isomist, Cooled Double Pass (5 °C) | Cyclonic Double Pass |
↵Notes: 1Stabilization of sample matrix in sample introduction and stabilization in plasma without fast pump flow rate to speed the uptake of the sample into the sample introduction.
↵2 Rinse time is based on washout of the highest concentration of elements to eliminate carryover. Sodium and phosphorous buffer samples required longer rinse time than pure water samples.
↵3 Per manufacturer's specifications.