Performance Parameters
RMM | |||
---|---|---|---|
Trade name | BactiFlow ALS | BacT/ALERT | AKuScreen |
Company | AES Chemunex | bioMérieux | Celsis |
Principle | Fluorescent labelling | Growth based | ATP Bioluminescence |
Detection system | Flow cytometry | CO2/Colour change | RLU/luminosity |
Detection type | Qualitative | Qualitative | Qualitative |
Particular | |||
Limit of detection | 1 cell per g+ | 1 cell per sample | 1 cell per g+ |
Sample volume | pre-incubation step (1 in 100 dilution) then 50 μL samples | Up to 10 mL per 40 mL vial | pre-incubation step (1 in 100 dilution) then 50 μL samples |
Capacity | ∼200/day | ∼50/day | ∼1000/day |
25/h | module dependant (∼240) | 120/h (164/run) | |
Time at which results are read** | 24 h | 24–72 h | 18 h |
Ease of use | |||
Amount of manipulation required | Medium | Easy | Medium |
Cost | |||
Hardware* | £90K | £30K | £67K |
Consumables per test | £4 | £4/vial | £4 |