TABLE XV

Levels of Extracted Elements, Thermoplastic Elastomer

Extraction DetailsAmount of Element Extracted, μg/g (ppm)4
SampleSolventVolume (mL)Temp. (°C)TimeWeight (g)AlBBaCaFeMgNaPSSiZn
12pH 3150404 days15.03333333330.603
pH 9150404 days15.0333333NA5NA50.600.103
E/W150404 days15.03333331.35.66.833
2pH 21001211 h3.533333333333
3pH 2.5150554 days15.033333333333
pH 9.5150554 days15.0333333NA5NA50.300.603
4pH 3150404 days15.03330.60330.80330.103
pH 9150404 days15.0333333NA5NA530.103
E/W150404 days15.03330.50331.911333
5pH 3150404 days15.03330.80333331.43
pH 9150404 days15.03330.6033NA5NA531.03
E/W150404 days15.03330.50332.23.63113
6pH 3501211 h1.233333333333
pH 9501211 h1.2333333NA5NA5333
E/W50404 days1.230.2533339.23333
  • Notes: 1The solvent types denoted with the pH were aqueous solutions prepared at that pH. Solvents at pH 3 and below were KCl/HCl mixtures. Solvents at pH above 3 were dilute phosphate buffers. W = purified water. E/W = 40/60 (v/v) ethanol/water.

  • 2 This material was gamma-irradiated.

  • 3 — = not present in the sample in reportable quantities.

  • 4 Elements that were not present in the extracts at reportable levels, in addition to those listed in this table, included Ag, As, Be, Bi, Cd, Co, Cr, Cu, Ge, Li, Mn, Mo, Ni, Pb, Sb, Se, Sn, Sr, Ti, V, and Zr.

  • 5 NA = not applicable. As Na and P were components of the buffer used in these solvents, extractable quantities of these elements could not be ascertained