Operating Conditions, ICP-MS System for Elemental Analysis
Parameter | Setting |
---|---|
Forward Power | 1300 watts (7500 a), 1500 watts (7500 c) |
Acquire Integration Time | 0.10 seconds per point |
Integration Mode | Auto |
Replicates | 1 |
Points per Peak | 6 |
Rinse Time | 180 s |
Rinse Rate | 0.5 rps |
Uptake Time | 35 s |
Uptake Rate | 0.5 rps |
Stabilization Time | 20 s |
Analysis Pump Rate | 0.1 rps |
Sample Introduction | Polypropylene Spray Chamber, Platinum Injector |
Nebulizer | Cross Flow |
Nebulizer Flow rate | 1.1 L/min |
Other Settings | Determined by tune results |