TABLE III

HVLD Precision—Differences among Test Days and between Bag Sources

MethodSensitivity set pointColumn A WAL-made bagsColumn B Novo Nordisk–made bagsColumn C WAL-made versus Novo Nordisk–made bags
Mean voltage Day-to-day agreementMean voltage Day-to-day agreementMean voltage agreement across test days
1Lower± 1 σ± 1 σ± 1 σ
Upper± 2 σ± 1 σ± 2 σ
2Lower± 2 σ± 2 σ± 1 σ
Upper± 3 σ± 2 σ± 1 σ
3Lower± 2 σ± 2 σ± 1 σ
Upper± 1 σ± 1 σ± 1 σ
4Lower± 2 σ± 1 σ± 1 σ
Upper± 1 σ± 1 σ± 1 σ
5Lower± 1 σ± 1 σ± 1 σ
Upper± 1 σ± 1 σ± 1 σ
6Lower± 1 σ± 1 σ± 2 σ
Upper± 1 σ± 1 σ± 1 σ
  • Note: σ = standard deviation