HVLD Test Methods
Test Method | Package orientation | HVLD test parameters | HVLD acceptance criteria | ||
---|---|---|---|---|---|
Metal clip (none, up, down) | Test direction (Septum or bag first) | High voltage set point (kV) | Sensitivity (midrange set point) | Ref Min to Ref Max limits (V) | |
1 | None | Septum first | 12.5 | 280 | 2.47–3.43 |
2 | None | Bag first | 12.5 | 260 | 2.42–3.50 |
3 | Up | Septum first | 12.5 | 255 | 2.34–3.18 |
4 | Up | Bag first | 12.5 | 245 | 2.42–3.26 |
5 | Down | Septum first | 12.5 | 260 | 2.31–3.33 |
6 | Down | Bag first | 12.5 | 245 | 2.40–3.30 |