TABLE I

HVLD Test Methods

Test MethodPackage orientationHVLD test parametersHVLD acceptance criteria
Metal clip (none, up, down)Test direction (Septum or bag first)High voltage set point (kV)Sensitivity (midrange set point)Ref Min to Ref Max limits (V)
1NoneSeptum first12.52802.47–3.43
2NoneBag first12.52602.42–3.50
3UpSeptum first12.52552.34–3.18
4UpBag first12.52452.42–3.26
5DownSeptum first12.52602.31–3.33
6DownBag first12.52452.40–3.30