TABLE III

HVLD of Poorly Sealed Bags

Bag IDHeat Seal SettingVisible Liquid Leakage (Y/N)HVLD MethodClip (None, Up, Down)Test Direction (Septum or Bag First)Final Analysis1 (Test Run Count)
34-11Manufactured sealN1NoneSeptumPass (4 of 4)
2NoneBagPass (4 of 4)
35-13.5Y1NoneSeptumFail
2NoneBagFail
35-23.5Y1NoneSeptumFail
2NoneBagFail
35-34.0N1NoneSeptumFail
2NoneBagFail
35-44.0Y1NoneSeptumFail
2NoneBagFail
35-54.5Y1NoneSeptumFail (1 of 2)
2NoneBagFail (1 of 2)
35-64.5N1NoneSeptumFail (1 of 2)
2NoneBagFail (1 of 2)
35-75.0N1NoneSeptumPass
2NoneBagPass
35-85.0N1NoneSeptumPass
2NoneBagPass
35-98.0N1NoneSeptumPass
2NoneBagPass
35-108.0N1NoneSeptumPass
2NoneBagPass