TABLE V

HVLD Test Methods

Test methodClip (none, up, or down)Test direction (septum or bag first)High voltage set point (kV)Sensitivity lower limit set pointSensitivity midrange set pointSensitivity upper limit set pointReference minimum limit (V)Reference maximum limit (V)
1NoneSeptum12.52752802852.473.43
2NoneBag12.52552602652.423.50
3UpSeptum12.52502552602.343.18
4UpBag12.52402452502.423.26
5DownSeptum12.52552602652.313.33
6DownBag12.52402452502.403.30