HVLD Test Methods
Test method | Clip (none, up, or down) | Test direction (septum or bag first) | High voltage set point (kV) | Sensitivity lower limit set point | Sensitivity midrange set point | Sensitivity upper limit set point | Reference minimum limit (V) | Reference maximum limit (V) |
---|---|---|---|---|---|---|---|---|
1 | None | Septum | 12.5 | 275 | 280 | 285 | 2.47 | 3.43 |
2 | None | Bag | 12.5 | 255 | 260 | 265 | 2.42 | 3.50 |
3 | Up | Septum | 12.5 | 250 | 255 | 260 | 2.34 | 3.18 |
4 | Up | Bag | 12.5 | 240 | 245 | 250 | 2.42 | 3.26 |
5 | Down | Septum | 12.5 | 255 | 260 | 265 | 2.31 | 3.33 |
6 | Down | Bag | 12.5 | 240 | 245 | 250 | 2.40 | 3.30 |